Compact Compressor NebulizerMODEL NE-C30INSTRUCTION MANUALENGLISH 35ESPAÑOLPORTUGUÊSCaution: Federal (USA) law restricts this device to sale by or on
PREPARING THE NEBULIZER FOR USE10Operating InstructionsWARNINGClean and disinfect the nebulizer kit, mouthpiece, and optionalnosepiece or masks before
PREPARING THE NEBULIZER FOR USE114. Lift up the baffle and remove the baffle from the medication cup.CAUTIONDo not alter the baffle, the nozzle in the
PREPARING THE NEBULIZER FOR USE8. Attach the desired inhalation accessory.NOTE: Mouthpiece and NosepiecePlace the cap on the medication cup when using
ATTACHING THE AIR TUBE131. Push the Air Plug on one end of the Air Tube onto the Air Connector on the front side of the compressor (A). 2. Push the A
USING THE DEVICE14WARNINGDo not use the device where the device may be exposed to flammable gas or vapors.WARNINGDo not use or store the device in hum
Press the power switch again to stop nebulization.NOTE: The green light goes off.WARNINGDo not cover the compressor with a blanket, towel, or any othe
5. Press the power switch. The compressor turns off and nebulization stops.6. Disconnect the Air Tube from the Nebulizer Kit. Hold the Air Plug and ge
Following the cleaning instructions after each use will prevent any remainingmedication in the bottle from drying resulting in the device not nebulizi
DAILY DISINFECTING18You can disinfect daily by soaking the parts in a vinegar solution or using acommercially available medical disinfectant. If your
DAILY DISINFECTING197. Submerge the parts in the white vinegar and distilled water solution for 30 minutes. 8. Remove the parts and discard the soluti
2CONTENTSBefore Using the DeviceIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHANGING THE AIR FILTER20Change the air filter every 60 days even if the air filter does not appear dirty.If the air filter appears dirty, or if water
CARING FOR THE DEVICE21To keep your device in the best condition and protect the unit from damagefollow these directions:CAUTIONDo not subject the com
CARING FOR THE DEVICE22STORING THE DEVICE 1. Put the nebulizer kit and the inhalation accessory (mouthpiece, nosepiece or mask) in a dry, sealed bag.
SOLUTIONCAUSEPROBLEMTROUBLESHOOTING GUIDE23No power on unit whenthe power switch is on. Turn the power switch off.Plug the power plug into anelectrica
LIMITED WARRANTIESYour Omron NE-C30 CompAir®Elite Nebulizer Compressor, excluding the nebulizer kit, inhalationaccessories, air tube and air filters,
FCC STATEMENT25Note:POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)This product has been tested and found to comply with the limits for
Model: NE-C30Type: Compressor nebulizerPower Supply: 12VDC 1A (Compressor) 100V-240VAC 50-60Hz (AC adapter)Power Consumption: 12 wattsNebulizer Rate:
NOTES27NE-C30-IM-FINAL 10/20/06 2:27 PM Page 26
Manufactured By OmronMade in JapanDistributed by:Omron Healthcare, Inc.1200 Lakeside DriveBannockburn, Illinois 60015www.omronhealthcare.comCopyright
3Thank you for purchasing the OMRON®NE-C30 Compact Compressor Nebulizer.INTRODUCTIONFill in for future reference.DATE PURCHASED: __________________SER
4SAFETY INFORMATIONTo assure the correct use of the product basic safety measures should always befollowed including the warnings and cautions listed
IMPORTANT SAFETY NOTES5OPERATING THE DEVICE (continued)Do not insert any object into the compressor.Make sure the air filter is clean. If the air filt
6IMPORTANT SAFETY NOTESRISK OF ELECTRICAL SHOCKDo not connect the battery connector located on the bottom of the main unit to the rechargeable battery
7IMPORTANT SAFETY NOTESMAINTENANCE AND STORAGEKeep the device out of the reach of unsupervised infants and children. The device may contain small part
Main Unit:CompressorComponents:Nebulizer Kit35KNOW YOUR UNITAir PlugAir PlugAir ConnectorAir TubeNebulizer KitHolderAir Filter Cover (*Air Filter Insi
9KNOW YOUR UNITComponents:Optional Accessories:User Registration CardAir Tubewith attached Air PlugsInstruction ManualAdult MaskNosepieceCarrying BagA